Chinese Journal of Hernia and Abdominal Wall Surgery(Electronic Edition) ›› 2025, Vol. 19 ›› Issue (04): 482-484. doi: 10.3877/cma.j.issn.1674-392X.2025.04.023

• Specialty Equipment Management • Previous Articles    

Establish the standard operating procedures in the management application of laparoscopic instruments for hernia and abdominal wall surgery

Yanjie Xu1, Jing Liu1, Chenghong Xiang1, Miaomiao Geng1, Jinfeng Guo2, Wei Liu1,()   

  1. 1Operation Room, Beijing Chaoyang Hospital, Capital Medical University, Beijing 100020, China
    2Operation Room, Chinese Academy of Medical Sciences Plastic Surgery Hospital, Beijing 100144, China
  • Received:2024-02-05 Online:2025-08-18 Published:2025-09-01
  • Contact: Wei Liu

Abstract:

Objective

To explore the application of standard operating procedures (SOP) in the management of laparoscopic instruments for hernia and abdominal wall surgery.

Methods

From April 2023 to October 2023, 360 sets of laparoscopic instruments for hernia and abdominal wall surgery in Beijing Chaoyang Hospital affiliated to Capital Medical University were selected and randomly divided into study group and control group. SOP were used to clean and manage laparoscopic instruments in the study group. The control group was cleaned according to the technical operation specification for endoscopic cleaning and disinfection 2023 edition, and managed according to conventional methods. The qualified rate of cleaning, the integrity rate of equipment, the normal rate of mechanical function and the correct rate of assembly were compared between the two groups.

Results

The qualified rate of cleaning in the study group was higher than that in the control group, the difference was statistically significant (P<0.001). The integrity rate, normal rate of mechanical function and correct assembly rate of the study group were higher than those of the control group, and the differences were statistically significant (P<0.05).

Conclusion

The establishment of SOP can significantly improve the cleaning effect of external hernia endoscopic instruments and avoid the possible risk of infection, ensure the quality of laparoscopic instruments and reduce the occurrence of errors, improve the assembly accuracy of endoscopic instruments, so as to effectively ensure the medical safety of patients.

Key words: Standard operating procedures, Laparoscopic instruments, Cleaning, Effect